•Positive or Negative Pressure Isolator. / mini clean room (not convertible)
•Work area and work surface built in AISI 304 stainless steel. The work surface is endowed with a vain for easy cleaning and collection of 4000 ml liquids with drain valve for cleaning.
•Front glass with butyl/neoprene/EPDM gloves, and ocassional opening allowing entry of large materials.
•Recycled air is filtered by an absolute HEPA H14 filter (efficiency 99.995 % MPPS EN1822) and is sent to the work area at a speed of 0.40 m/s ± 0.05 m/s.
• Expulsion of 100% of the air, without recirculation.
•The air expelled to the outside is filtered by an absolute filter HEPA H14 (efficiency 99.995% MPPS REGULATION EN1822) independent dedicated for this purpose.
•Work area maintains laminar flow conditions with iso 3 air purity (according to ISO 14644).
•Microprocessor controlled cockpit with updatable FLASH memory and external display for easy display of cabinet control parameters (speeds, flow rates, plenum pressure, ejection barriers, etc...).
•Rubber terminal seal to ensure watertightness.
•SIDE SAS with UV and two doors interlocked.
•The isolator is equipped with an easy-to-use touch panel with safety key for white/ultraviolet lighting control, internal electrical outlet, gas valve and parameters.
•Problem detection alarms in the drive system and negative/possitive pressure protection barrier and door opening.
•Exclusive fluorescent and UV lighting.
•Working voltage 220 V. 50 Hz. Maximum current consumption 6 A. (approx. 1300 w.).
•As standard the cab Includes U.V. light and LED lighting; Electrical socket of 110-220 volts depending on the customer's need.
•The UV light has a timer.
Optional Accessories:
- •Gas intake with electronically controlled valve.
- •Vacuum socket.
- •Peroxide decontamination connector Hydrogen and/or formalin.
- •Informatization with Integrated Monitor and Connections USB, HDMI, VGA.
Advantages:
- •ISO 3 Isolator / mini clean room according to ISO 14644.(Class A - GMP)
- •LCD display. All functions visible on the screen (laminarity, barriers, temperature, humidity, filter filling, etc...)
- •Remote control. All the functions of the cab can be controlled six meters away, so that the user can be protected in case of emergency.
- •Ergonomically tested, thanks to the noise level that is less than 60 dBA, below the level required by the 65 dBA standard.
- •Monotorization and adjustment of the filling of the filters minimizing the cost of replacing them.